10. Rapid Alert – Spurious Injection RHOPHYLAC 300 µg (Batch # P100707423)

Rapid Alert

SPURIOUS INJECTION RHOPHYLAC 300 µg (BATCH NO. P100707423)
DRAP Alert NoNo I/S/03-26-10
Action Date11 March, 2026
Target Audience1. Regulatory Field Force of DRAP and Provincial Drug Control departments.
2. Healthcare Professionals
3. Pharmacies and medical stores
Problem StatementDrug Testing Laboratory (DTL), Punjab vide Test Report No. 01-105012931/DTL dated 06-02-2026 has declared the following sample as Spurious under Section 3(zb)(ii) of the Drugs Act, 1976. The product is purported to be manufactured by M/s CSL Behring AG, Switzerland and imported/distributed by M/s Hakimsons (Impex) (Pvt.) Ltd., Karachi; however, verification from the stated importer confirms that the recovered batch does not belong to them and was not legally imported.
The sample also failed Sterility Test (Membrane Filtration Method) and was additionally categorized as Substandard on sterility grounds and Misbranded with respect to labeling.
S#Product NameBatch No.Manufacturer (as per label)Test Result
1Injection Rhophylac 300 µg (Human Anti-D (Rh) Immunoglobulin) Prefilled Syringe 2 ml (1500 IU / 300 µg)P100707423Purported to be manufactured by: M/s CSL Behring AG, Switzerland Imported/Distributed by: M/s Hakimsons (Impex) (Pvt.) Ltd., KarachiDeclared Spurious, Misbranded & Substandard (Sterility Failure)
Risk StatementThe product is a Human Anti-D (Rh) Immunoglobulin injection indicated for prevention of Rh isoimmunization in Rh-negative individuals. Laboratory testing revealed microbial growth in sterility testing, indicating contamination and rendering the product unsafe for parenteral administration. Additionally, the batch has been confirmed as falsified and not legally imported by the authorized distributor. Administration of this contaminated and spurious biological product may lead to serious bloodstream infections, therapeutic failure in prevention of Rh sensitization, and life-threatening complications. The public health risk is assessed as high, particularly for pregnant women and immunocompromised patients.
Action InitiatedThe Regulatory Field Force of DRAP and Provincial Drug Control Departments has been directed to conduct surveillance activities throughout the supply chain to confiscate the falsified products.
Advice for Healthcare ProfessionalsDRAP requests healthcare professionals to have increased vigilance within the supply chains of institutions/pharmacies/healthcare facilities likely to be affected by above mentioned product. Any adverse events or quality problems experienced with the use of these products should be reported to  National or Provincial pharmacovigilance centers using Adverse Event Reporting Form or online through this link.  Further information on reporting problems to DRAP is available on this link.
Advice for ConsumersConsumers should not use these products and should contact their physician or healthcare provider(s) if they have experienced any problem related to taking or using the above mentioned products and should report the incident to the Drug Regulatory Authority of Pakistan / National Pharmacovigilance Centre. All therapeutic goods must be obtained from licensed pharmacies and other authorized/licensed retail outlets. The authenticity and condition of products should be carefully checked. Seek advice from your pharmacists or other healthcare professionals in case of any doubt.